VIG 20/2 DUAL CHANNEL FID
Now available for rent, the VIG 20/2 dual channel head total hydrocarbon analyzer designed for high accuracy, sensitivity and stability.
The unit uses two flame ionization detectors (FIDs) for continuous measurement of two independent sample streams.
All components that come in contact with the sample through analysis are maintained in a temperature-controlled oven to prevent condensation, and to provide repeatable, reliable performance in the analysis of a wide variety of hydrocarbon concentrations in gaseous mixtures or in ambient air.
Features:
Automatic start-up / ignition
Choice of Hydrogen or Hydrogen/Helium fuel
Heated sample pump heads
Two stage sample filters with exchangeable sintered stainless steel elements
Applications:
Automatic start-up / ignition
Choice of Hydrogen or Hydrogen/Helium fuel
Heated sample pump heads
Two stage sample filters with exchangeable sintered stainless steel elements
Ambient Temperature | From 50°F to 120°F |
Voltage Outputs | 0-10VDC (standard) |
Display | Graphic backlit 240W x 64H pixels high contrast wide viewing angle |
Operation Requirements | Fuel - UHP Hydrogen @ 18psi incoming pressure (standard) |
NULL | Zero Calibration Gas - UHP zero grade air or nitrogen @ 9psi incoming pressure |
Measuring Method | 2 Oven Heated Flame Ionization Detectors (FIDs) |
Measurement Range/Standard Ranges | (4 Ranges per amplifier 2 amplifiers per analyzer) 0-10 0-100 0-1000 0-10000ppm (Lower detection limit 0.01ppm) or |
Zero & Span Noise | Less than 0.2% of full scale |
Zero & Span Drift | +/- 1% full scale per 24 hours |
Linearity/Repeatability/Stability | Within 1% of full scale through all ranges |
Response Time | Within 5 seconds 90% of final reading |
Dimension
US | Metric | |
Dimensions | 19 x 16.75 x 24 in | - |
Weight
US | Metric | |
Weight | 35 lbs | - |