RAE Systems ppbRAE 3000 PID
RAE Systems ppbRAE 3000 PID
RAE Systems ppbRAE 3000 PID
RAE Systems ppbRAE 3000 PID
RAE Systems ppbRAE 3000 PID
RAE Systems

RAE Systems ppbRAE 3000 PID

Pine Item #51242

The RAE ppbRAE 3000 is the most advanced handheld volatile organic compound (VOC) monitor on the market. Its photoionization detector (PID) extended range of 1 ppb to 10,000 ppm makes it an ideal instrument for applications from HazMat/Homeland Security and industrial hygiene to Indoor Air Quality and military applications.

Features:

Big graphic display

Designed for simple service

Field-interchangeable battery pack

Integrated flash light

Integrates RAE Systems Correction Factors list for more than 200 compounds

Multi-language support

Proven PID technology

Real-time wireless

Rugged housing

User-friendly screens including dataplot chart view

Applications:

Decontamination

Environmental Cleanups and Remediation

Indoor Air Quality

Hazmat Reponse

Worker Exposure

Venue Protection

Urban Search and Rescue

Shut downs and turnarounds

Quantity must be 1 or more
Translation missing: en.products.product.equipment_available
Request a Quote

Specifications

Technical Specifications

Sensors Photoionization sensor with standard 10.6 eV or optional 9.8 eV or 11.7 eV lamps
Battery Rechargeable external field-replaceable Lithium-Ion battery pack. Alkaline battery adapter
Operating hours 16 hours of operation (12 hours with alkaline battery)
Graphic Display 4 lines 28 x 43 mm with LED backlight for enhanced display readability
Direct Readout Instantaneous reading - VOCs as ppm by volume; High values; STEL and TWA; Battery and shutdown voltage; Date time temperature
Alarms 95 dB (at 30 cm) buzzer and flashing red LED to indicate exceeded preset limits High: 3 beeps and flashes per second; Low: 2 beeps and flash
EMI/RFI Highly resistant to EMI/RFI Compliant with EMC Directive 89/336/EEC
IP Rating IP67 unit off and without flexible probe IP65 unit running
Datalogging Standard 6 months at one-minute intervals
Calibration Two-point or three-point calibration for zero and span. Calibration memory for 8 calibration gases alarm limits span values and calibratio
Sampling Pump Internal integrated flow rate at 500 cc/mn. Sample from 100' (30m) horizontally and vertically
Low Flow Alarm Auto pump shutoff at low-flow condition
Communication Download data and upload instrument set-up from PC through charging cradle or optional Bluetooth Wireless data transmission through built-i
Frequency 902 to 928 MHz (license-free) 2.400 to 2.4835 GHz (license-free) 433 MHz 869 MHz
RF Range Up to 500' (900 MHz 433 Mhz 869 Mhz) extendable with RAELink3 Repeater to 2 miles
Hazard Area Approval US and Canada: UL cUL Classified as Intrinsically Safe. Approval for use in Class I Division I Groups A B C D. Europe: ATEX II 1G EEx
Temperature -4° to 113° F (-20° to 50° C)
Humidity 0% to 95% relative humidity (non-condensing)
Attachments Durable black rubber boot
SENSOR SPECIFICATIONS VOCs
Range 0 to 9999 ppb Resolution - 1 ppb
Range 10 to 99 ppm Resolution - 0.01 ppm
Range 100 - 999.9 ppm Resolution - 0.1 ppm
Range - 1000 - 9999 ppm Resolution - 1 ppm
Response Time T9 3s
Dimensions US: 10 L x 3.0\ W x 2.5\""" H\""" Metric: 25.5cm x 7.6cm x 6.4cm
Weight US: 26oz Metric: 738g